NovaSeq 6000 Sequencing System specifications

Powerful performance at hand

Scalable, high-throughput sequencing with tunable output

NovaSeq 6000 system render

Output per flow cell for various read lengthsa

Read length SP S1 S2 S4
1 × 35 bp N/A N/A N/A 280–350 Gb
2 × 50 bp 65–80 Gb 134–167 Gb 333–417 Gb N/A
2 × 100 bp 134–167 Gb 266–333 Gb 667–833 Gb 1600–2000 Gb
2 × 150 bp 200–250 Gb 400–500 Gb 1000–1250 Gb 2400–3000 Gb
2 × 250 bpb 325-400 Gb N/A N/A N/A

a. Output specifications based on VRÕæÈ˲ÊƱ PhiX control library at supported cluster densities.ÌýNovaSeqÌý6000 system is capable of single flowÌýcell runs or dual flow cell runs. Dual flow cell run delivers twice the output listed above.Ìý​

b. Maximum read length.

Estimated sample throughput for key applicationsa

Application SP S1 S2 S4
Human genomes per run ~4 ~8 ~20 ~48
Exomes per run ~40 ~80 ~200 ~500
Transcriptomes per run ~32 ~64 ~164 ~400

a. All sample throughputs are estimates and are based on dual flow cell runs. Human Genomes assumes > 120 Gb of data per sample to achieve 30× genome coverage. Exome assumes ~8 Gb/100×. Transcriptomes assume ≥ 50M reads. Throughput may vary based on library preparation kit used.

Reads passing filter per flow cella

Read type SP S1 S2 S4
Single-end reads 650–800²Ñ 1.3–1.6µþ 3.3–4.1µþ 8–10µþ
Paired-end reads 1.3–1.6µþ 2.6–3.2µþ 6.6–8.2µþ 16–20µþ

a. All sample throughputs are estimates and are based on dual flow cell runs. Human Genomes assumes > 120 Gb of data per sample to achieve 30× genome coverage. Exome assumes ~8 Gb/100×. Transcriptomes assume ≥ 50M reads. Throughput may vary based on library preparation kit used.

Quality scoresa

Read length SP S1 S2 S4
1 × 35 bp N/A N/A N/A ≥ 90% of bases higher than Q30
2 × 50 bp ≥ 90% of bases higher than Q30 N/A
2 × 100 bp ≥ 85% of bases higher than Q30
2 × 150 bp ≥ 85% of bases higher than Q30
2 × 250 bp ≥ 75% of bases higher than Q30 N/A N/A N/A

a. Quality score based on NovaSeq 6000 Reagent Kits v1.5 using an VRÕæÈ˲ÊƱ PhiX control library.

Run timea

Read length SP S1 S2 S4
1 × 35 bp N/A N/A N/A ~14 hr
2 × 50 bp ~13 hr ~13 hr ~16 hr N/A
2 × 100 bp ~19 hr ~19 hr ~25 hr ~36 hr
2 × 150 bp ~25 hr ~25 hr ~36 hr ~44 hr
2 × 250 bp ~38 hr N/A N/A N/A

a. Run time based on NovaSeq 6000 Reagent Kits v1.5 using an VRÕæÈ˲ÊƱ PhiX control library while running two flow cells simultaneously.

Instrument specifications

NovaSeq 6000 system specifications

NovaSeq 6000 instrument specifications

  • Instrument configuration
    Computer and touch screen display Installation setup and accessories data collection and analysis software
  • Instrument control computer
    Note: Computer configurations will be upgraded regularly; contact your local account manager for current configuration
    Base Unit: Portwell WADE-8022 with Intel i7 4700EQ CPU
    Memory: 2 × 8 GB DDR3L SODIMM
    Hard Drive: None
    Solid-State Drive: 256 GB mSATA
    Operating System: Windows 10
  • Operating environment
    For Indoor Use Only
    Temperature: 19°C-25°C (22°C ±3°C), <2°C change per hour
    Humidity: 20%-80% relative humidity, non-condensing
    Altitude: Below 2000 meters (6500 feet)
    Ventilation: Maximum of 8530 BTU/h and average 6000 BTU/h
  • Laser
    Wavelengths
    532 nm, 660 nm, 780 nm, 790 nm
  • Dimensions with monitor
    W × D × H
    80.0 cm × 94.5 cm × 165.6 cm
    (31.5 in × 37.2 in × 65.2 in)
  • Weight
    481 kg (1059 lb)

    Includes 3.5 kg (7.8 lb) for leak tray and 0.9 kg (2 lb) for keyboard and mouse

  • Crated Dimensions
    W × D × H
    92 cm × 120 cm × 118 cm
  • Crated Weight
    628 kg (1385 lb)
  • Power requirements
    200–240 VAC 50/60Hz, 16A, single phase, 2500 W VRÕæÈ˲ÊƱ provides a region-specific uninterruptible power supply
  • Network connection
    Dedicated 1 Gb connection between the instrument and data management system. Connect directly or through network
  • Bandwidth for network connection
    200 Mb/s/instrument for internal network uploads
    200 Mb/s/instrument for BaseSpace Sequence Hub uploads
    5 Mb/s/instrument for Instrument Operational Data uploads

Key technologies